Physiol Meas. 2016 May;37(5):649-60. doi: 10.1088/0967-3334/37/5/649. Epub 2016 Apr 19.

The study of in vivo quantification of aluminum (Al) in human bone with a compact DD generator-based neutron activation analysis (NAA) system.

Byrne P, Mostafaei F, Liu Y, Blake SP, Koltick D, Nie LH.
  • School of Health Sciences, Purdue University, West Lafayette, IN 47907, USA.

 

Abstract

The feasibility and methodology of using a compact DD generator-based neutron activation analysis system to measure aluminum in hand bone has been investigated. Monte Carlo simulations were used to simulate the moderator, reflector, and shielding assembly and to estimate the radiation dose. A high purity germanium (HPGe) detector was used to detect the Al gamma ray signals. The minimum detectable limit (MDL) was found to be 11.13 μg g(-1) dry bone (ppm). An additional HPGe detector would improve the MDL by a factor of 1.4, to 7.9 ppm. The equivalent dose delivered to the irradiated hand was calculated by Monte Carlo to be 11.9 mSv. In vivo bone aluminum measurement with the DD generator was found to be feasible among general population with an acceptable dose to the subject.

PMID: 27093035

 